Products

200mm DAC system
Key words:
Classification:
200mm DAC system
- Product Description
- Technical parameters
- More introduction
-
- Commodity name: 200mm DAC system
- Commodity ID: 1091282555085934592
Product Introduction:
Equipped with dynamic axial compression column200mmPilot scale-up system, usingH-TREEPatented distribution design, better separation effect. The preparation column and the column loading system adopt the integrated design, and the column loading process is very simple,10μmThe packing can achieve almost the same high column efficiency as the analysis column, can be used online after the preparation of the column is installed without disassembly, and the axial dynamic pressure is always maintained inside the column bed, which can maintain the stability of the column bed and the column efficiency for a long time, and the single sample loading amount can be more achieved.20gAbout, this product is very suitable for annual production100kgCustomers of the following pure products.
Product Features:
1) It is convenient to install the column, and the operator can install the column by himself.
2) The column performance is maintained for a long time and the column has a long service life.
3) AdoptH-TREEPatented distribution design, better separation effect
4) Dynamic axial compression columns can be customized according to user needs.
5) The injection mode can be selected by a separate injection pump or injection valve.
6) The maximum and minimum pressure can be adjusted, the pump will stop immediately when the maximum pressure is higher than the maximum pressure, and the pump will be delayed when the minimum pressure is lower than the minimum pressure.2minAutomatic pump stop.
-
Product parameters :
Chromatographic column
Column inner diameter:
200mm
Column tube length:
650mm (standard configuration)
Column tube material:
316L stainless steel
Inner wall finish:
Ra≤0.4μm
Piston material:
316L stainless steel
Sieve plate material:
316L stainless steel, pore size: 3-5μm
Maximum withstand voltage:
10MPa
Air pressure requirements:
≥0.8MPa
Gas demand:
≥3m 3
Pump
Infusion method:
Double plunger reciprocating motion
pump head:
3000mL stainless steel pump head
Maximum working pressure:
10MPa
Flow range:
(0.001~3000)mL/min
Flow accuracy:
±1%
Flow repeatability:
RSD≤1.0%
Operating mode:
isocratic or gradient
Collection device:
Automatic or manual collection devices can be configured according to user needs
Detector
Wavelength range:
( 190-800 )nm
Light source:
Deuterium lamp ( standard configuration ); tungsten lamp ( replaceable )
Wavelength Accuracy:
±0.75nm
Wavelength repeatability:
0.2nm
Static baseline noise:
≤ 0.5 × 10 -5 AU
Static baseline drift:
≤ 1.5 × 10 -4 AU/h
Flow cell:
10L flow cell (standard configuration)
-
Application areas:
1) drug purification:Synthetic drugs, fermented drugs, biological pharmaceuticals, etc.
2) purification of natural products
3) purification of fine chemical intermediates
Commonly used consumables(Optional):
1) BalsealSealing ringX225920 3LPump high pressure sealing ring
2) 3Lift pump low pressure sealing ring
3) 3One-way valve for lift pump
4) UV3000detector deuterium lamp
5) 200DACPiston lower seal ringF4
6) 200DACSealing ring of piston sieve plate
7) 200DACSealing ring of sieve plate at bottom of column
8) P185Axial sealing ring(200AHigh-pressure sealing ring for axial use)
9) 210*5.33CladdingORing
10) FV158X2.5PTFE-coated seal ring
11) Sealing ring of sieve plate at bottom of column
Product inquiry
Note: Please leave your email and our professionals will contact you as soon as possible!